廠牌:三菱化學分析科技
供應商:南北貿易有限公司 日東精工分析科技(三菱化學分析科技) 台灣總代理
聯絡電話:04-22994269
快速詢價 |
商品規格 | |
商品名稱:
|
總硫分析儀 TCL-100型 |
分析方法:
|
高溫氧化分解庫倫滴定法 |
燃燒:
|
矽管燃燒 |
高溫爐:
|
1100oC |
樣品種類:
|
液體及氣體樣品 |
檢測方法:
|
氧化-還原電位(以電極檢測) |
滴定控制:
|
自動電解電流控制 |
測定範圍:
|
微量~1000ppm |
測定時間:
|
約10分 |
操作控制:
|
PC配合軟體程式Data Base |
所需氣體:
|
Ar 99.8%以上, O2 99.7%以上 |
符合標準:
|
氯:ASTM D5808, ASTM D4929…等 |
選購品 | |
|
|
ASC-150L (Liquid Auto sampler) 適用於液體樣本 |
|
|
|
CRI-100V (Constant Rate Injector)
|
|
|
|
GI-100 氣體注射器 適用於氣體樣本 |
|
|
TCL-100 | |
The field-proven coulometry system allows quick measurement of all chlorine contained in a liquid or gaseous sample in different concentrations ranging form ppm to percent levels. Our analyzer allows fully automatic measurement of all chlorine in waste oil or reclaimed oil to reduce petroleum products, petrochemical products, and dioxin. |
|
Features
Windows® XP is employed for easier operation and advanced functions.
|
TCL-100 Specifications | |
|
|
Analysis method: | Oxidative decomposition, coulometric titration |
|
|
Combustion: | Silica tube combustion |
|
|
Combustion furnace temperature: |
1,100oC at he maximum |
|
|
Samples to analyze: | Liquid and gaseous samples |
|
|
Detection method: | Oxidation-reduction potential (electric potential detection by electrodes) |
|
|
Detecting electrode: | [Chlorine analyzer]: Silver electrode [Sulfur analyzer]: Alloy electrode |
|
|
Titration control: | Automatic electrolytic current control |
|
|
Gas to use: | Argon: Purity greater than 99.98%, 300±100kPa Oxygen: Purity greater than 99.7%, 300±100kPa |
|
|
Measurement range: [Chlorine analyzer] |
0.1 to 50μg (CI), 0.5 to 1,000ppm (Samples burnable below 1,000oC with no difficulty) |
|
|
Measurement accuracy: [Chlorine analyzer] |
Standard samples: CV of 5% or below for 1μg of chlorine When 2,4-dinitrochlorobenzene is measured |
|
|
Measurement range: [Sulfur analyzer] |
0.02 to 20μg (S), 0.1 to 1,.000ppm (Samples burnable below 1,000oC with no difficulty) |
|
|
Measurement accuracy: [Sulfur analyzer] |
Standard samples: CV of 5% or below for 1μg of sulfur When dibutyl disulfide is measured |
|
|
Measurement time: | approx. 10min./sample (content of 2μg at measurement) |
|
|
Amount of samples to measure: |
Liquid: 5 to 200μl*1, Gas: 5to 10ml*2 |
|
|
Dimensions, weight: | Main unit: approx. 500(W)×400(D)×500(H)mm, 35kg Detector: approx.220(W)×370(D)×415(H)mm, 8kg |
|
|
*1 : The constant rate injector can be used with a liquid sample. In this case, up to 200μl of a sample can be injected. *2 : The gas injector is required for a gaseous sample.
|
|
Options Depending on the samples to analyzed, optional products can be selected to configure a desired system. |
||||||||||||||||||
|
||||||||||||||||||
Liquid Auto Sampler
|
||||||||||||||||||
|
||||||||||||||||||
Constant Rate Injector
model CRI-100V CRI-100V can inject non-aqueous samples at a constant rate to a microsyringe. Syringe: 50μl, 100μl, 250μl (gas-tight syringes) Injection rate: 0.2 to 1.3μl/sec. (variable) Dimensions, weight: approx. 250(W)×150(D)×288(H)mm, 7kg |
||||||||||||||||||
|
||||||||||||||||||
Gas Injector
model GI-100 GI-100 is used to analyze gas samples. Samples to inject: Gaseous samples Standard liquid sample: Max. 20μl (Gastight syringe) Standard gaseous sample: Max. 10μl (Gastight syringe) Gas to use: Argon (purity greater than 99.98%) Dimensions, weight: approx. 120(W)×250(D)×235(H)mm, 4kg |
||||||||||||||||||
|
||||||||||||||||||