商品名稱:
|
微量總硫分析儀 TS-100型 |
測定方法:
|
高溫氧化裂解紫外螢光法 |
樣品導入:
|
樣品舟自動導入 |
樣品取量:
|
固體1~30mg,液體5~50μl,氣體1-10ml |
測量範圍:
|
極微量~10000ppm |
分析時間:
|
3~10分 |
操作控制:
|
PC配合軟體程式Data Base |
所需氣體:
|
Ar 99.98%以上, O2 99.97%以上 |
適用對象:
|
粗氣油、汽油、燈油、輕油、原油、重油、潤滑油、發動機燃料、液化石油汽BTX(苯、甲苯、二甲苯)…等各產品,石油化學產品。 例:Naphtha, Gas Oil, Kerosene, Gasoline Lubricating Oil, Fuel Oil, LPG, PBT...etc. |
選購測定:
|
總氮同時分析化學發光法 |
符合標準:
|
硫:ASTM D5453, ASTM D6667… |
選購品 | |
|
|
ASC-150L (Liquid Auto sampler) 適用於液體樣本 |
|
|
|
ASC-120S (Solid Auto Sampler)
|
|
|
|
ND-100 (Nitrogen Detecor)
|
|
|
|
CRI-100H (Constant Rate Injector)
|
|
|
|
GI-100 氣體注射器 適用於氣體樣本 |
|
|
TS-100 | |
Trace amounts of sulfur can be measured rapidly by a completely dry technique with a sulfur detector employing an ultraviolet fluorescent method. The use of database software for data processing allows all aspects of measurement to be managed with a database for easy retrieval of data.
|
|
Features
Highly Sensitive Analysis of Sulfur and NitrogenThe TS-100 Trace Sulfur Analyzer is a completely dry analyzer that uses an oxidation decomposition and ultraviolet fluorescence method. The sulfur detector of the TS-100 offers a high level of sensitivity that enables stable measurement of sulfur content down to the ppb level. Addition of an optional nitrogen detector using the field-proven oxidation decomposition and chemiluminescence method enables simultaneous detection of both sulfur and nitrogen. This analysis is used in the certified methods indicated below. Oxidation decomposition and ultraviolet fluorescence ASTM-D5453 Oxidation decomposition and chemiluminescence JIS-K2609 ASTM-D4629, D5176, D5762 and D6069 Outstanding Ease of Operation and Improved Functions Using Windows®95/98 The system can be operated easily simply by using a mouse. All measurement conditions are contained in the database, allowing samples to be measured simply by selecting the optimum conditions. In addition, a full range of help functions are available that can be used as a supplement to the instruction manual.
High-Speed Analysis Through the Use of Automatic Boat Controller (ABC) System Configuration Corresponding to the Purpose of Use |
TS-100 Specifications 《 Sulfur measurement (Standard) 》 | |
|
|
Samples : | Solid, liquid and gaseous samples (Naphtha, Gas oil, Kerosene, Gasoline, Lubricating oil, Fuel oil, LPG, PBT...etc.) |
|
|
Analytical method : | Oxidative pyrolysis-ultraviolet fluorescence method |
|
|
Sample introduction method: |
Automatically introduced using sample boat (using automatic boat controller) |
|
|
Sample size: | Solid samples: 1-30mg, liquid samples: 5-50μl*1, gaseous samples: 1-10μl*2 |
|
|
Measurement range: | 0.05-10,000 ppm |
|
|
Measurement time: | 3-10 min./ measurement |
|
|
Interface: | Balance input (RS232C) |
|
|
Furnace temperature: | Max. 1,100oC |
|
|
Gas requirements : | Argon (purity:99.98% or more, 290-390 kPa) Oxygen( purity:99.97% or more, 290-390 kPa) *3 |
|
|
Dimensions, weight: | Main Unit: approx. 560(W)×400(D)×435(H)mm, 17kg ABC : approx. 440(W)×250(D)×180(H)mm, 11kg |
|
|
*1 :The amount of sample injected can be increased up to 250μl by using the optional constant rate injector (CRI-100H) when measuring liquid samples. |
|
Options | |
|
|
Liquid Auto Sampler model ASC-150L Liquid samples are measured, injected and rinsed automatically using a microsyringe. Up to 50 sample vials can be placed in the sampler, and 50 samples can be processed automatically, including sampling, measurement, data processing and report generation. Furthermore, this sampler is used in combination with the Automatic Boat Controller provided as standard equipment. |
|
|
|
Solid Auto Sampler
model ASC-120S Up to 20 solid samples can be measured automatically. The main unit contains a built-in automatic boat controller, and samples weighed in the quartz boats are measured according to predetermined combustion conditions. |
|
|
|
Nitrogen Detector
model ND-100 Connecting this nitrogen detector to the TS-100 system allows measurement of total nitrogen in non-aqueous systems and solid samples. |
|
|
|
Constant Rate Injector
model CRI-100H (for ND-100*) Liquid samples are injected at a constant rate using a microsyringe. Even samples that burn with difficulty can be measured with high accuracy by controlling the injection arte. |
|
|
|
Gas Injector
model GI-100 This system is used for analysis of sulfur and nitrogen in gaseous samples. |
|
|
|