M7500流變儀是專為對複雜樣品進行簡單測試而設計的一款高溫、超高壓、低剪切、全自動數字流變儀。本儀器專利的測量機構設計取消了昂貴和易損的寶石軸承,可以進行寬範圍的測量。由於該儀器採用了多種獨特的結構設計,使其便於維護並大大簡化了操作流程。自動化的壓力、速度和溫度控制,使實驗結果更加精確和一致。
基於微軟數據庫作為友好的用戶界面,它能夠允許用戶在全球範圍內共享測試結果。
優勢:
特點:
規格:
尺寸: |
22" x 12" x 24" (壓力室) 12" x 25" x 15.5" (控制單元) |
重量: |
250 lbs |
材質: |
不銹鋼 |
樣品體積: |
132 ml |
溫度範圍: |
環境溫度 (20 °F 使用冷卻器) ~ 500 °F |
速度範圍: |
0.01 ~600 rpm 連續 |
黏度範圍: |
0.5 ~ 5,000,000 Cp |
力矩範圍: |
7 μN.m ~10 mN.m |
壓力範圍: |
大氣壓 ~ 20,000 psi 或 30,000 psi |
剪切速率: |
0.0082 ~ 1020 S-1 |
剪切應力: |
2 ~ 1,600 dyn/cm2 |
重複性: |
±1% 全量程 |
精度: |
0.3% 全量程 |
電壓等級: |
120 VAC 或240 VAC (配置變壓器) |
冷卻介質: |
自來水或冷卻水 |
壓縮空氣: |
120 psi |
電腦配置: |
MS Windows 9X/2000/XP/Vista |
Description:
Patented design prevents sample contamination
The Grace Instrument M7500 Ultra HPHT Rheometer is a coaxial cylinder, rotational, high pressure, and high temperature rheometer. It is engineered to measure various rheological properties of fluids (including API HPHT tests) under a range of pressures and temperatures, up to 30,000 psi and 600°F. An optional cement module is also available for testing the rheology of cements (see tab "Cement Rheometer", below.)
The Grace Instrument M7500 Ultra HPHT Rheometer employs a unique, patented design which entirely prevents contact between sample fluid within the main test chamber and pressurization fluid, which is injected into interlocking isolated chambers above the test chamber. No other rheometer on the market today can make this claim.
Optional components add multiple test functions while saving time, money, and laboratory space
Hardware modules are available which add specific functionality to the M7500, including cement testing, pressure/volume testing, linear swell testing, and more. These modules take the place of stand-alone equipment and can be integrated into the standard M7500 training program, saving training time and consolidating test times. This feature alone enables the M7500 Ultra HPHT Rheometer to provide an excellent ROI.
Features:
Comparison between Grace M7500 and Leading Competitor | |||
---|---|---|---|
Grace M7500 | Competitor | ||
Bearings | Patented and patent pending design eliminates jewel bearings. Use ball bearings. Easy to replace. Very economic. |
V Shape jewel bearings. Fragile, easy to break, difficult to install, very expensive. | |
Mixing with Pressurization Fluid | Patented and patent pending design prevents tested sample to mix with pressurization fluid as long as tested sample is compressed less than 23% of volume. | Tested sample will mix with pressurization fluid when tested sample is compressed more than 2.5% of sample volume. | |
Speed Range | Standard range 0.01 to 600 RPM. | 3 to 600 Rpm continuously. | |
Electronics | Almost all electronic components are from third party reputable companies with excellent quality tracking records. Easy to replace in the future. | Majority of the electronics are special designed PCB by fann Instrument. Easy to break down. Very difficult to replace and re-calibrate. Very expensive to maintain. | |
Operation | Patented and patent pending design very easy to operate. Test turn around time is about one quarter of fann 75. No need for re-calibration before every test. | Need special training to perform testing. Very difficult to operate due to extremely complex mechanical designs. Need calibration before every single test. Very difficult to calibrate too. | |
Software | State of the art PC software with Database engine, real-time charting, various 2D & 3D charts, and unlimited chart overlapping. Database management of test results, searchable by test name, additives, etc. | Limited functions designed almost 10 years ago. | |
Calibration | Automatic shear stress calibration. | Requires user to go through some complicated steps and mechanical adjustment before able to start calibration. | |
Weight & Dimensions | Separate cell tower and control cabinet. Smaller foot print, height and weight. Movable by two personnel | One integrated unit. Can only be moved by 4 people. Dangerous to move around due to its whole weight. |
Specifications:
Dimensions: 22" height x 12" width x 24" depth (Tower) 12" height x 25" width x 15.5" depth (Cab) |
Weight: 250 lbs. |
Construction: Stainless Steel |
U.S. Patent: 7,287,416 and 7,412,877 |
Sample Size: 132 ml |
Temperature Range: Ambient (20°F with chiller) to 600°F |
Speed Range: 0.01 to 600 rpm continuous |
Viscosity: 0.5 to 5,000,000 Centipose |
Torque: 7μN.m to 10 mN.m |
Pressure Range: Atm. to 30,000 psi |
Shear Rate: 0.0082 to 1020 S-1 |
Shear Stress: 2 to 10,000 dyn/cm2 |
Repeatability: ±1% of full scale range or better |
Resolution: 0.3% of full scale range or better |
Voltage: 120 VAC or 240 VAC (with transformer) |
Coolant Supply: Tap water or chiller |
Compressed Air: 120 psi |