•量測範圍(MIR):0~10ppm、0~100ppm、0~1,000ppm、0~10,000ppm、0~100,000ppm 5個量測範圍甲烷濃度,可手動及自動切換。
•偵測極限:0.1ppm甲烷
•樣品氣體輸入上最大壓力偏差:±50mbar
•T90(無樣品管路):約2秒
•分析溫度:100~200℃
•分析壓力:100mbar
•樣品氣體流量(無樣品管路):約90 L/hr(=1500 mL/min)
•樣品氣體分析量:<15 L/hr(=250 mL/min)
•燃料氣體:H2、H2/He
•燃料消耗:H2(約20 mL/min)、H2/He(約80 mL/min)
•燃燒空氣:內涵燃燒空氣產生器
•歸零氣體:『內』-內部啟動碳過濾器、『外』-快速接頭或樣品進口
•警報:探測器火焰、分析溫度、樣本氣體通量
•分析儀耗電量:250VA(初始約450VA)
•操作環境溫度:+5~+45℃
•重量:約19 kg
(Reference Quantity: Full Range Deflection = FRD)
•5 Measurement Ranges
0-10 ppm Propane
0-100 ppm Propane
0-1.000 ppm Propane
0-10.000 ppm Propane
0-100.000 ppm Propane
•Linearity:+/- 2 % MBE (up to Range 4)
•Oxygen Cross Sensitivity:< +/- 2 % FRD (with He/H2 Fuel)
•Limit of Detection:< 2 % FRD
•Zero Driftt:< 2 % FRD/week
•Span Drift:< 4 % FRD/week
•Response Time90:approx. 1-2 s
•Sample Flow Rate:approx. 90 l/h
•Analytical Temperature:up to 200°C
•Signal Output:0-10 V and/or 0/4-20 mA
•Status Contacts:Concentration-Alarm、Range Identification、Flame Out Indication、Sample Flow Alarm
•Dimensions:360 x 200 x 500 mm (WxHxD)
•Weight:approx. 20 kg
•Ambient Temperature:+5 ... +35 °C
•Main Power:230 V / 50 Hz or 115 V / 60(50) Hz / 450 Watts