Vessel |
Working volume (Liter) |
3L |
5L |
7L |
10L |
15L |
20L |
Total volume (Liter) |
3.8 |
6.8 |
8.9 |
12.5 |
18 |
24.4 |
Material |
Borosilicate glass / 316L stainless steel for headplate and all fittings |
Control Unit |
Control panel |
10.4" Colorful touch-screen interface |
Communication port |
Remote control through Ethernet (SCADA) |
Data export through USB port |
RS-485 port for system extension |
Program storage |
Up to 59,994 process programs |
Log data storage |
Up to 10 process monitoring data files |
Cabinet material |
ABS front panel and painted iron housing |
Dimension |
Footprint: W400 x D500 mm (15.75" x 19.69") |
Height = 740 mm (29.13") |
Rated voltage |
110V or 220V ; 50/60 Hz |
Aeration |
Inlet gas flow-meter |
0, 1-10 LPM |
0, 2-20 LPM |
0, 4-50 LPM |
Sparger |
Orifice ring |
Baffle |
Removable 316L stainless steel baffles |
Temperature |
Heating |
Heating blanket |
Cooling |
Cooling coil |
Range |
5°C (41°F) above coolant up to 60°C (140°F) |
Resolution |
0.1°C |
Probe |
Platinum RTD probe (Pt 100) |
Control mode |
Programmable 15 steps PID control |
Agitation |
Drive |
Removable top brushless motor |
Speed range |
Two agitation motor types: |
a. For fermentation and cell culture: 30 - 1,200 rpm |
b. For shear-force sensitive cell culture line: 5 - 300 rpm (optional) |
Resolution |
1 rpm |
Impeller |
Two impeller types: |
a. For fermentation: 3 pcs of adjustable Rushton-type impeller |
b. For shear-force sensitive cell culture: 2 pcs of adjustable Pitched-blade impeller (Optional) |
Note: Customized impellers are available |
Control mode |
Programmable 15 steps PID control |
pH |
Range |
2 -14 pH |
Resolution |
0.01 pH |
Probe |
Gel-filled electrode: Autoclavable |
Control mode |
PID, One point control with adjustable dead band |
DO |
Range |
0 - 200% |
Resolution |
0.1% |
Probe |
Polarographic DO sensor; Autoclavable |
Control mode |
Cascade function to response to |
a. Increase or decrease agitation speed |
b. Oxygen enrichment module (optional) |
c. Start substrate feeding program |
ORP |
Measurement range |
-/+ 2000mV |
Resolution |
1 mV |
Probe |
Gel-filled electrode: Autoclavable (optional) |
Foam |
Probe |
316L stainless steel detector with insulated Teflon tube; |
On/Off control |
Peristaltic pump |
Pump number |
4 Built-in pumps; One external pump (optional, RS-485 interface with relay output) |
Motor type |
Precise stepping motor; minimum speed is 1 rpm |
Speed range |
0 - 65 rpm |
Resolution |
1 rpm |
Control mode |
Programmable 15 steps feeding control; |
Pump can be assigned for Acid, Base, Antifoam and Substrate |
Exhaust |
Device type |
316L stainless steel condenser |
Utility requirement |
Power Supply |
100-120V 50/60Hz or 210-230V 50/60Hz with electrical safety cutoff switch |
Water |
2 Bar maximum (29 psi) |
Air |
1 Bar maximum, must be dry, oil-free and filtered (14.5 psi) |
Autoclave |