廠牌:TURNER DESIGNS/U.S.A.
供應商:建成科學儀器股份有限公司
聯絡電話:02-29103818
快速詢價 |
Turner Designs公司最新推出 Trilogy™ 實驗室用螢光計. 採用先進光學模組化設計, 為一多功能的實驗室設備.
使用snap-in應用模組可輕易進行螢光度, 吸光度, 濁度三種測量. Trilogy™ 具備觸摸顯示面板, 使得操作及校正過程更為直接, 同時可儲存18組校正曲線, 每組曲線可為5點校正.
葉綠素a測定
藍綠藻測定
水質監測
螢光追蹤研究
食品處理
研究探討
Fluorescence Modules: 螢光應用光學模組
Turbidity Module: 濁度光學模組
Absorbance Module & Filter Paddles: 吸光度光學模組
An Overview
Numerous Applications exist in:
· Chlorophyll a Measurement
· Cyanobacteria Measurement
· Water Quality Monitoring
· Fluorescent Tracer Studies
· Food Processing
· Research Studies
Features of the Trilogy™
‧Snap-in Application Modules enable you to select the applications you need today while being able to easily add in the future
‧Application Modules available for several fluorescent applications, turbidity measurements, and absorbance readings
‧Simple user interface based on a touchscreen with menu selections
‧Flexible sample compartment accommodates 10 x 10 mm square plastic cuvettes, 12 x 75 mm round tubes, and 12 x 35 mm round vials
‧One to five point calibration with up to 18 calibrations stored
‧Adjustable solid secondary standard provides easy calibration verification
‧EPA 445 approved acidification and non-acidification chlorophyll a measurement calculations
‧Spreadsheet Interface Software provides real time display and logging of data to an Excel spreadsheet
Instrument purchase includes: adjustable solid standard, 10 ea 10 x 10mm plastic cuvettes and 12 x 35 mm glass vials, 50 well test tube holder, RS-232 cable, and power supply
Fluorescence Modules:
Turbidity Module:
Absorbance Module & Filter Paddles:
Fluorescence Module Performance
Minimum Detection Limit: 0.02 ug/L Chlorophyll a,
0.02 ppb Rhodamine WT
Dynamic Range: 0-4000 ppb RWT
Linearity: 0.99R2
Turbidity Module Performance
Minimum Detection Limit: Contact us
Dynamic Range: Contact us
Linearity: Contact us
Absorbance Module Performance
Photometric Measuring Range: 0.0 - 4.0A
Photometric Accuracy: +/- 0.7%
Photometric Precision: <0.5% at 1A
General Specifications
Sample Adaptors: Accommodates 10 x 10 mm square plastic cuvettes, 12 x 75 mm round tubes, and 12 x 35 mm round vials.
Readout: Direct Concentration (ug/L, ppb, etc.) or Raw Fluorescence
Calibration: One to Five point calibration with up to 18 calibrations stored
Light Source & Detector: Light Emitting Diode and Photodiode
Blank: Reads and subtracts blank
Data Output: 100% ASCII format through a 9 pin RS-232 serial cable at 9600 baud
PC Operating System (optional if connected to PC): Windows 98â or later
Power: 100 to 240 VAC Universal Power Supply included Output 12VDC 0.84A Max
Dimensions: 12.92"D x 10.44"W x 8.42"H (32.82 cm D x 26.52 cm W x 21.39 cm H)
Weight: 8.1 lbs (3.65 kg)
Operating Temperature: 60 - 105 °F (15-40 °C)
Humidity: 75% RH maximum
Warranty: One year
Approvals: CE