Data Cable: 1 meter, for use with data port
Data Port: RS-232 (serial), for PC interface
Display: Graphical LCD with backlight
Flow Measurement Cell: Integrated (all-in-one design)
Flow Measurement Style: Single, Continuous or Burst, with averaging function userselectable from 1 to 100
Flow Measurement Type: Volumetric (primary by nature), Standardized
Gas Compatibility: Noncorrosive, noncondensing, noncombustible gases, less than 70% humidity
Humidity (Ambient): 0-70%, noncondensing
Inlet and Outlet Fittings: 1 1/2” Swagelok® compression tube fitting
Overpressure Release Fitting: 1/2” Swagelok® compression tube fitting
Storage Temperature: 0-70o C
Power: External power module Input: 100-240 VAC, 1.6A (max), 50-60 Hz Output: 12 VDC, 3.0A
DryCal 1020 Pressure (Operating): 10 - 19.5 PSI
DryCal 1500 Pressure (Operating): 14-45 PSI
DryCal 1020 Pressure Transducer: 0-20 PSI (absolute), accuracy +0.05% FS (typical), +0.10% (max)
DryCal 1500 Pressure Transducer: 0-50 PSI (absolute), accuracy +0.05% FS (typical), +0.10% (max)
Purge Fittings: 1/4” Swagelok® compression tube fittings
RoHS Compliant: Yes
Temperature: 15-30o C
Valve Operation Air supply: 80-100 PSI (not provided) Fitting: 1/4th” Swagelok® compression
Warranty: 1 year (battery 6 months)